| DSI PhysioTel® Transmitter Specifications |
PhysioTel® TA Transmitter Specifications
PhysioTel® PA Transmitter Specifications
PhysioTel® EA, CA, ETA, and CTA Transmitter Specifications
PhysioTel® Multiplus Transmitter Specifications
| PhysioTel® TA Transmitter Specifications |
Model |
Weight
(gm) |
Volume
(cc) |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
TA-F20 |
3.8 |
1.75 |
20
grams |
33 x 33 x 14
cm |
6 |
TA-F40 |
7.25 |
3.5 |
175
grams |
42 x 42 x 18
cm |
12 |
TA-D70 |
40 |
25 |
2.5
kilograms |
1 x 1 x 1
meter |
11 |
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel® TA transmitters in animals smaller than 20 grams. Larger cage sizes also can be accommodated with special configurations.
| Parameters Measured |
Temperature and Activity |
| Outer Material |
Silicone Elastomer |
| Temperature Operating Range |
34 to 41 degrees Celsius |
| Shelf Life |
Battery - 1 year
Sterility - 1 year |
| ON-Off Mechanism |
Magnetically Actuated |
| Receiver Compatibility |
Any PhysioTel® RLA, RPC-1 or RMC-1 |
| |
|
Temperature Measurement Specifications |
| Initial Accuracy |
0.1 degree Celsius |
| Resolution |
0.01 degree Celsius |
| Max Drift Over |
0.05 degree Celsius first two weeks after turn on |
| Warranted Battery Life |
0.05 degree Celsius per six months thereafter |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
| PhysioTel® PA Transmitter Specifications |
Model |
Weight
(gm) |
Volume
(cc) |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
PA-C10 |
1.4 |
1.1 |
17
grams |
33 x 33 x 14
cm |
1.5 |
PA-C40 |
9.0 |
4.5 |
175
grams |
42 x 42 x 18
cm |
4 |
PA-D70 |
37 |
25 |
2.5
kilograms |
1 x 1 x 1
meter |
4 |
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted the PhysioTel® PA-C10 transmitter in mice smaller than 17 grams. Larger cage sizes also can be accommodated with special configurations. Minimum animal weight may vary based upon placement location, surgical skill and strain of animal used.
| Parameters Measured |
Pressure and Activity |
| Outer Material |
Silicone Elastomer |
| Temperature Operating Range |
34 to 41 degrees Celsius |
| Shelf Life |
Battery - 1 year
Sterility - 1 year |
| ON-Off Mechanism |
Magnetically Actuated |
| Receiver Compatibility |
Any PhysioTel® RLA, RPC-1 or RMC-1 |
| |
|
Measurement Specifications |
| Pressure Range |
-20 to 300 mmHg |
| Ambient Pressure Range |
670 to 800 mmHg |
| Maximum Altitude |
625 meters |
| |
|
Pressure Accuracy |
| Initial Accuracy |
±3 mmHg |
| |
|
Pressure Drift |
| PA-C40 |
First month < 5 mmHg
Second month < 4 mmHg
Thereafter < 3 mmHg per month |
| PA-C10, PA-D70 |
< 2 mmHg per month nominal |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
| PhysioTel® EA, CA, ETA and CTA Transmitter Specifications |
Model |
Weight
(gm) |
Volume
(cc) |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
EA-F20 |
3.9 |
1.9 |
20
grams |
33 x 33 x 14
cm |
4 |
ETA-F20 |
3.9 |
1.9 |
20
grams |
33 x 33 x 14
cm |
4 |
CA-F40 |
7.0 |
3.0 |
175
grams |
42 x 42 x 18
cm |
6 |
CTA-F40 |
7.0 |
3.0 |
175
grams |
42 x 42 x 18
cm |
6 |
CTA-D70 |
30.0 |
20.0 |
2.5
kilograms |
1 x 1 x 1
meter |
4 |
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers may have implanted PhysioTel® EA and ETA transmitters in animals smaller than 20 grams. Larger cage sizes also can be accommodated with special configurations.
| Parameters Measured |
Biopotential, Temperature, Activity |
| Outer Material |
Silicone Elastomer |
| Temperature Operating Range |
34 to 41 degrees Celsius |
| Shelf Life |
Battery - 1 year
Sterility - 1 year |
| ON-Off Mechanism |
Magnetically Actuated |
| Receiver Compatibility |
Any PhysioTel® RLA, RPC-1 or RMC-1 |
| |
|
Temperature Measurement Specifications |
| Initial Accuracy |
0.1 degree Celsius |
| Resolution |
0.01 degree Celsius |
| |
|
Max Drift Over Warranted Battery Life |
| F20 and F40 series |
0.1 degree Celsius first two weeks after turn on
0.05 degree Celsius per six months thereafter |
| D70 series |
0.05 degree Celsius first two weeks after turn on
0.05 degree Celsius per six months thereafter |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
| PhysioTel® Multiplus Transmitter Specifications |
Model |
Weight
(gm) |
Volume
(cc) |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
F20-EET |
3.9 |
1.9 |
20
grams |
33 x 33 x 14
cm |
1.5 |
F40-EET |
7 |
3.5 |
175
grams |
42 x 42 x 18
cm |
3 |
F40-TT |
7.5 |
3.5 |
175
grams |
42 x 42 x 18
cm |
4 |
F50-EEE |
11.5 |
5.5 |
175
grams |
42 x 42 x 18
cm |
2 |
C50-PXT |
11 |
6 |
175
grams |
42 x 42 x 18
cm |
2 |
D70-PCT |
49 |
33 |
2.5
kilograms |
1 x 1 x 1
meter |
3.5 |
D70-CCTP |
49 |
33 |
2.5
kilograms |
1 x 1 x 1
meter |
2.5 |
D70-PCTP |
49 |
33 |
2.5
kilograms |
1 x 1 x 1
meter |
2.5 |
D70-EEE |
37 |
25 |
2.5
kilograms |
1 x 1 x 1
meter |
2 |
4ET-S1
and
4ET-S2 |
12.8 |
8.8 |
200
grams |
42 x 42 x 18
cm |
3
(user replaceable battery) |
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel® Multiplus transmitters in animals smaller than 175 grams. Larger cage sizes also can be accommodated with special configurations.
| Parameters Measured |
Pressure, Biopotential, Temperature, Activity |
| Outer Material |
Silicone Elastomer |
| Temperature Operating Range |
34 to 41 degrees Celsius |
| Shelf Life |
Battery - 1 year
Sterility - 1 year |
| ON-Off Mechanism |
Magnetically Actuated |
| Receiver Compatibility |
Any PhysioTel® RLA, RPC-1 or RMC-1 |
| |
|
Measurement Specifications |
| Pressure Range |
-20 to 300 mmHg |
| Ambient Pressure Range |
670 to 800 mmHg |
| Maximum Altitude |
625 meters |
| |
|
Pressure Accuracy |
| Initial Accuracy |
±3 mmHg |
| Pressure Drift (C50-PXT) |
First month < 5 mmHg
Second month < 4 mmHg
Thereafter < 3 mmHg per month |
| Pressure Drift (D70 Series) |
<2 mmHg per month nominal |
| |
|
Temperature Accuracy |
| Initial Accuracy (D70-CCTP, D70-PCTP) |
0.35 degree |
| Initial Accuracy (Others) |
0.25 degree |
| Resolution |
0.05 degree Celsius |
| Temperature Drift (D70-CCTP, D70-PCTP) |
0.2 degree Celsius per month |
| Temperature Drift (F20-EET) |
0.15 degree Celsius per month |
| Temperature Drift (Others) |
First two weeks .1 degree Celsius
Thereafter .1 degree Celsius |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
Request more information on this product.
|