Data Sciences International
  • Home

DSI PhysioTel® Transmitter Specifications

PhysioTel® HD Transmitter Specifications
PhysioTel® TA Transmitter Specifications
PhysioTel® PA Transmitter Specifications
PhysioTel® EA, CA, ETA, and CTA Transmitter Specifications
PhysioTel® Multiplus Transmitter Specifications
PhysioTel® F40-TT Transmitter Specifications

PhysioTel® HD Transmitter Specifications

Model

Weight
(gm)

Volume
(cc)

Minimum Animal Weight

Maximum Cage Size*

Battery Life Warranted (Months)

HD-X11

2.2

1.4

19 grams
(sub-cu)
30 grams (IP)

33 x 33 x 14 cm

1

HD-S21

8

5.9

175
grams

42 x 42 x 18
cm

2

Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel HD transmitters in animals smaller than minimum animal weight specified. Larger cage sizes also can be accommodated with special configurations.

Parameters Measured Pressure, Biopotential, Temperature, Activity
Outer Material Silicone Elastomer (HD-S21)
Temperature Operating Range 34 to 41 degrees Celsius
Shelf Life Battery - 1 year
Sterility - 1 year
ON-Off Mechanism Magnetically Actuated
Receiver Compatibility

Any PhysioTel® RLA, RPC-1 or RMC-1

Pressure Measurement Specifications

Pressure Range -20 to 300 mmHg
Catheter Length Options HD-S21: 8, 10 or 15 cm
HD-X11: 5 cm
Ambient Pressure Range 670 to 800 mmHg
Maximum Altitude

625 meters

Pressure Accuracy

 
Initial Accuracy ± 3mmHg
Pressure Drift

<2 mmHg/month (maximum); -0.25mmHg/month (Average)
Frequency Response

HD-S21: 8cm catheter: >100Hz (Minimum Performance), 120Hz (Average Performance)†
HD-X11: 5cm catheter: >60Hz (Minimum Performance); 75Hz (Average Performance)

Biopotential Channel Specifications

HD-S21

HD-X11

Input Voltage Range ±5mV
± 2.5 mV
Channel Bandwidth  0.1-145 Hz
0.1 - 200 Hz


Temperature Accuracy

 
Initial Accuracy 0.1 degree Celsius
Resolution 0.05 degree Celsius
Temperature Drift

First two weeks ≤ .1 degree Celsius
Per 6 months after ≤ .1 degree Celsius

Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.

*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.

†Recommended catheter length for small animal LVP applications.

PhysioTel® TA Transmitter Specifications

Model

Weight
(gm)

Volume
(cc)

Minimum Animal Weight

Maximum Cage Size*

Battery Life Warranted (Months)

TA-F10

1.6

1.1

17 g (sub-cu)
20g (IP)

33 x 33 x 14
cm

6

TA-F40

7.25

3.5

175
grams

42 x 42 x 18
cm

12

TA-D70

40

25

2.5
kilograms

1 x 1 x 1
meter

11

Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel TA transmitters in animals smaller than the specified minimum animal weight. Larger cage sizes also can be accommodated with special configurations.

Parameters Measured Temperature and Activity
Outer Material Silicone Elastomer
Temperature Operating Range 34 to 41 degrees Celsius
Shelf Life Battery - 1 year
Sterility - 1 year
ON-Off Mechanism Magnetically Actuated
Receiver Compatibility Any PhysioTel® RLA, RPC-1 or RMC-1
   

Temperature Measurement Specifications

Initial Accuracy 0.15 degree Celsius (TA-F10), 0.1 degree Celsius (all other devices)
Resolution 0.05 degree Celsius (TA-F10), 0.01 degree Celsius (all other devices)
Max Drift Over Warranted Battery Life 0.05 degree Celsius first month after turn on (TA-F10)
0.05 degree Celsius first two weeks after turn on
(all other devices)
0.05 degree Celsius per six months thereafter
(all devices)

Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.

*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.

PhysioTel® PA Transmitter Specifications

Model

Weight
(gm)

Volume
(cc)

Minimum Animal Weight

Maximum Cage Size*

Battery Life Warranted (Months)

PA-C10

1.4

1.1

17 grams
(sub-cu)
27 grams (IP)

33 x 33 x 14
cm

1.5

PA-C40

9.0

4.5

175
grams

42 x 42 x 18
cm

4

PA-D70

37

25

2.5
kilograms

1 x 1 x 1
meter

4

Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted the PhysioTel PA-C10 transmitter in mice smaller than 17 grams. Larger cage sizes also can be accommodated with special configurations. Minimum animal weight may vary based upon placement location, surgical skill and strain of animal used.

Parameters Measured Pressure and Activity
Outer Material Silicone Elastomer
(Excluding the PA-C10)
Temperature Operating Range 34 to 41 degrees Celsius
Shelf Life Battery - 1 year
Sterility - 1 year
ON-Off Mechanism Magnetically Actuated
Receiver Compatibility Any PhysioTel® RLA, RPC-1 or RMC-1
   

Measurement Specifications

Pressure Range -20 to 300 mmHg
Ambient Pressure Range 670 to 800 mmHg
Maximum Altitude 625 meters
   

Pressure Accuracy

Initial Accuracy ±3 mmHg
   

Pressure Drift

PA-C10 < 2 mmHg/month (Max); -0.25mmHg/month (Average)
PA-C40 First month < 5 mmHg
Second month < 4 mmHg
Thereafter < 3 mmHg per month
PA-D70 < 2 mmHg per month nominal

Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.

*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.

 

PhysioTel® EA, CA, ETA and CTA Transmitter Specifications

Model

Weight
(gm)

Volume
(cc)

Minimum Animal Weight

Maximum Cage Size*

Battery Life Warranted (Months)

ETA-F10

1.6

1.1

17 grams

33 x 33 x 14
cm

2

EA-F20

3.9

1.9

20
grams

33 x 33 x 14
cm

4

ETA-F20

3.9

1.9

20
grams

33 x 33 x 14
cm

4

CA-F40

7.9

4.2

175
grams

42 x 42 x 18
cm

6

CTA-F40

8.0

4.2

175
grams

42 x 42 x 18
cm

6

CTA-D70

30.0

20.0

2.5
kilograms

1 x 1 x 1
meter

4

Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers may have implanted PhysioTel CA and CTA transmitters in animals smaller than 175 grams. Larger cage sizes also can be accommodated with special configurations.

Parameters Measured Biopotential, Temperature, Activity
Outer Material Silicone Elastomer
Temperature Operating Range 34 to 41 degrees Celsius
Shelf Life Battery - 1 year
Sterility - 1 year
ON-Off Mechanism Magnetically Actuated
Receiver Compatibility Any PhysioTel® RLA, RPC-1 or RMC-1


Biopotential Channel Specifications

ETA, EA Devices

CTA, CA Devices

Input Voltage Range ±2.5mV
±10mV
Channel Bandwidth  1-200 Hz
1 - 200 Hz


Temperature Measurement Specifications

Initial Accuracy 0.1 degree Celsius
Resolution 0.01 degree Celsius
ETA-F10 Initial Accuracy

0.25 degrees Celsius

ETA-F10 Resolution 0.05 degrees Celsius

Max Drift Over Warranted Battery Life

F20 and F40 series 0.1 degree Celsius first two weeks after turn on
0.05 degree Celsius per six months thereafter

D70 series 0.05 degree Celsius first two weeks after turn on
0.05 degree Celsius per six months thereafter

Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.

*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.

PhysioTel® Multiplus Transmitter Specifications

Model

Weight
(gm)

Volume
(cc)

Minimum Animal Weight

Maximum Cage Size*

Battery Life Warranted (Months)

F20-EET

3.9

1.9

20
grams

33 x 33 x 14
cm

1.5

F40-EET

7

3.7

175
grams

42 x 42 x 18
cm

3

F50-EEE

11.5

5.5

175
grams

42 x 42 x 18
cm

2

C50-PXT

11

6

175
grams

42 x 42 x 18
cm

2

D70-PCT

49

33

2.5
kilograms

1 x 1 x 1
meter

3.5

D70-CCTP

49

33

2.5
kilograms

1 x 1 x 1
meter

2.5

D70-PCTP

49

33

2.5
kilograms

1 x 1 x 1
meter

2.5

D70-PCTR

49
33
2.5 kilograms     
1 x 1 x 1 meter  
2.5

D70-EEE

37

25

2.5
kilograms

1 x 1 x 1
meter

2

4ET-S1
and
4ET-S2

12.8

8.8

200
grams

42 x 42 x 18
cm

3
(user replaceable battery)

Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel Multiplus transmitters in animals smaller than 175 grams. Larger cage sizes also can be accommodated with special configurations.

Parameters Measured Pressure, Biopotential, Temperature, Activity
Outer Material Silicone Elastomer
Temperature Operating Range 34 to 41 degrees Celsius
Shelf Life Battery - 1 year
Sterility - 1 year
ON-Off Mechanism Magnetically Actuated
Receiver Compatibility Any PhysioTel® RLA, RPC-1 or RMC-1
   

Measurement Specifications

Pressure Range -20 to 300 mmHg
Ambient Pressure Range 670 to 800 mmHg
Maximum Altitude 625 meters
   

Pressure Accuracy

Initial Accuracy ±3 mmHg
Pressure Drift (C50-PXT) First month < 5 mmHg
Second month < 4 mmHg
Thereafter < 3 mmHg per month
Pressure Drift (D70 Series) <2 mmHg per month nominal
   


Biopotential Channel Specifications

F20-EET, F40-EET, 4ET

D70-EEE, F50-EEE

Input Voltage Range ±1.25mV
±2.5mV
Channel Bandwidth  1-50 Hz
-100 Hz
 

C50-PXT

D70-CCTP, PCT, PCTP, PCTR

  ±5mV
±10mV
  1-100 Hz
1-100 Hz

Temperature Accuracy

Initial Accuracy (D70-CCTP, D70-PCTP) 0.35 degree Celsius
Initial Accuracy (Others) 0.25 degree Celsius
Resolution 0.05 degree Celsius
Temperature Drift (D70-CCTP, D70-PCTP) 0.2 degree Celsius per month
Temperature Drift (F20-EET) 0.15 degree Celsius per month
Temperature Drift (Others) First two weeks .1 degree Celsius
Thereafter .1 degree Celsius

Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.

*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.

Request more information on this product.

PhysioTel® F40-TT Transmitter Specifications

Model

Weight
(gm)

Volume
(cc)

Minimum Animal Weight

Maximum Cage Size*

Battery Life Warranted (Months)

F40-TT

7.5

3.5

175
grams

42 x 42 x 18
cm

4

 

Temperature

Temperature Operating Range 22-41 degrees Celsius
Initial Accuracy 0.30 degree Celsius from 34-40 degree Celsius
0.45 degree Celsius all other
Temperature Drift 0.1 degree Celsius per 6 months
Resolution 0.01 degree Celsius
   
Battery Voltage  
  Continuous monitoring via parameter
Probe  
  1.8mm diameter
13cm thermistor-tipped probes
   

Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.

*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.

Request more information on this product.