7700 Series Isolated/Defibrillation Protected ECG and General Purpose Probes
Features
- 3 Lead connection with an array of electrodes
- Provides galvanic isolation to the signal conditioner
- Protects signal conditioner from defibrillator discharges
- Built-in “Zero” and “Cal” buttons for ease in calibrating channel
The Isolated Defibrillation Protected ECG and General Purpose Probes (13-7758-IDP and 13-7758-GPP) are designed to conduct pre-clinical studies and investigation (clinical research) of physiological processes. The probes are suitable for intra cardiac, surface electrodes, Action Potential measurements or other applications where isolation is needed. Additionally, built-in Zero and Cal buttons provide for a simple, accurate calibration process. The probes interface specifically to the 7700 Series Universal Signal Conditioner, model 13-7715-58.
Technical Data Sheet
Model 13-7758-IDP and 13-7758-GPP
Isolated/Defibrillation Protected Probes |
| SPECIFICATIONS |
Electrical Specifications
Gain x10
Maximum Output Range ±2.5VDC
Input Leads RA, LA, RL (+, -, REF)
Dynamic Range ±250mV
Maximum input DC offset ±300mV
Input Impedance 50MΩ @ 10Hz when using fully shielded cable (ConMed FSR 397 with patient cable and lead wires).
CMRR -90dB @ 60Hz with 51kΩ, 0.047uF source imbalance
Input Risk Current < 10μA @ 265V RMS 60Hz, sink all leads to ground
Defibrillator Protection 360Joules delivered at 5000Volts
Isolation 1500V RMS for 1 minute, isolated circuits to chassis
Noise <30μV p/p @ 1kHz
Bandwidth 0.05 to 3.5kHz
Calibration 1mV (ECG), 50 mV (GPP)
Physical Specifications
Height 1.2 inches
Width 3.2 inches
Length 5.9 inches
Weight 0.5 pounds
Environmental Specifications
Temperature
Operating 0 to 40° C
Storage 20 to 70° C
Humidity RH, non-condensing
Operating 10% to 95%
Storage 5% to 95%
Cooling Convection only
Altitude 2000 meters maximum |
| ORDERING INFORMATION |
Part Number Description
13-7758-IDP ACQ-7700 Universal Signal Conditioner
Accessories
P02141 ECG Cable Kit (Contains the following 5
items)
P02143-2 Grabber Lead Wire, set of 3
P02143-3 Banana Lead Wire, set of 3
P02143-4 Banana to CLIP Adapter, 3 each
P02143-5 0.80 Receptacle Lead Wire, set of 3
Safety
EN 60601-1: 1990, Medical Electrical Equipment, Part 1
General Requirements for Safety Including Amendments 1 and 2: 1995
EN 60601-2-25: 1993, Medical Electrical Equipment, Part 2
Particular requirements for the safety of electrocardiographs. Except for clause 51.103 Indication of inoperable ELECTROCARDIOGRAPH.
Emissions
EN 55011: 1991, Conducted and Radiated Emissions, Class A
EN 61000-3-2: 1995, Harmonics, Class A
EN 61000-3-3: 1995, Flicker
Immunity
EN 61000-6-2 and EN 60601-1-2
EN 61000-4-2: 1995, Electrostatic Discharge, 2,4,8 kV
EN 61000-4-3: 1998, Radiated EM field, AM modulation, 10 V/m
ENV 50204: 1995 Radiated EM Field, Pulse modulation,10 V/m
EN 61000-4-4: 1995, Fast Transient / Burst, 1,2 kV
EN 61000-4-5: 1995, Surge Immunity 0.5, 1, 2 kV
EN 61000-4-6: 1996, Conducted RF Disturbances, 10 V/rms
EN 61000-4-11: 1994, Voltage Dips, 95%/10ms, 70%/.1s, 40%/.5s, 95%/5s |